Abstract

Imaging the electrical output activity of biological cells is important to gain an understanding of how cell networks process information. This has implications for the understanding of brain processing, such as that performed by the retina in encoding the visual scene. The performance and electrical quality of a state-of-the-art high-density 519-microelectrode array, that recorded simultaneously from hundreds of live retinal output cells (ganglion cells) is reported on. The fabrication process for these devices has been optimised and their electrical characteristics examined. The electrode arrays typically exhibit an impedance of ∼200 kΩ at 1 kHz and the RMS noise of the whole recording system is 7 μV with a signal to noise ratio of 20:1. With a view to direct stimulation of retinal ganglion cells, a low impedance Z=300 kΩ iridium oxide interface capable of delivering large currents Q cap=4 mC/cm 2 to cells was also developed.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.