Abstract

We evaluated the performance of matrix-assisted laser desorption ionization–time of flight (MALDI-ToF) for identification of Bcc species compared with that of recA sequencing. MALDI-ToF was able of identifying 100% of Bcc isolates at the genus level, but 23.1% of Bcc isolates tested were not correctly identified at the species level. The misidentification occurred most frequently with Burkholderia contaminans (100%) and B. cepacia (33.3%).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.