Abstract

Previously we proposed a new depth of interaction (DOI) encoding method and proved that it worked successfully with four-layered Gd/sub 2/SiO/sub 5/ crystals for a small animal positron emission tomography (PET) detector. We are now planning a small animal PET scanner, jPET-RD (for Rodents with DOI detectors), which has both high resolution and high sensitivity. Scintillator for the scanner will be Lu/sub 2(1-x)/Y/sub 2x/SiO/sub 5/ (LYSO). In this work, we evaluated the DOI detector composed of four layers of 12/spl times/12 LYSO (Lu: 98%, Y: 2%) crystal array by irradiating 511 keV uniform gamma rays. For crystal identification, the new encoding method was used. The size of each crystal is 1.44 mm/spl times/1.44 mm/spl times/4.5 mm. The crystal block was coupled to a 256-channel flat panel position sensitive photomultiplier tube, which has 16/spl times/16 multi anodes at intervals of 3.04 mm. In this measurement, all crystals are expressed on a single two-dimensional position histogram without overlapping as we expected. Energy resolution of all events is 21.8 % and time resolution of all events is 0.69 ns in FWHM. The energy resolutions of the first, second, third, and fourth are 11.6%, 12.3%, 13.3%, and 12.5% and the time resolutions are 0.60 ns, 0.59 ns, 0.60 ns, and 0.66 ns, respectively.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.