Abstract
The determination of zinc ion (1–60 ng ml −1) by anodic square-wave stripping voltammetry on an anion-exchange perfluorinated polymer Tosflex mercury film electrodes (TMFE) was evaluated. The detection limit was 0.1 ng ml −1 Zn(II). The effect of various organic compounds (gelatin, albumin, starch, camphor, humic acid, Triton X-100, sodium dodecyl sulfate (SDS) and cetyltrimethylammonium bromide (CTAB)) is explored. The results indicate that due to the size-exclusion and ion-exchange properties of Tosflex film, the TMFE is considerably more resistant to organic interference than the bare mercury film electrode.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.