Abstract

The double balloon technique for aortic valvuloplasty using a combined brachial and femoral artery approach with readily available angioplasty equipment is described. Online hemodynamic assessment of the patient during the procedure was provided by continuous, reflective spectorphotometric monitoring of the pulmonary artery mixed venous oxygen saturation and the arterial pressure in the femoral sheath. Successful valvuloplasty was achieved in 10 patients with the transaortic peak-to-peak systolic gradient reduced from 77.6 +/- 28 to 27.8 +/- 15 mmHg (P less than 0.001), the mean gradient from 62.0 +/- 24 to 26.0 +/- 15 mmHg (P less than 0.0007), and the aortic valve area increased from 0.56 +/- 0.3 to 1.03 +/- 0.6 cm2 (P less than 0.05). No significant complications were encountered. The double balloon technique for aortic valvuloplasty is an acceptable alternative method to the single balloon technique.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.