Abstract

Patent foramen ovale (PFO) is a congenital, small tunnel-like connection between the right and left atria that usually closes spontaneously after birth. However, frequently (in up to 35% of the normal population) it persists into adulthood. It is associated with ‘paradoxical’ embolism from the venous to the arterial system and may result in stroke or peripheral embolism. One prophylactic treatment option is transcatheter closure of the PFO. Currently available closure devices extend into both atria and therefore occasionally cause complications, such as thrombus formation or erosion of adjacent structures. The Coherex FlatStent™ is a flat, self-expanding stent that is designed to be positioned within the PFO tunnel. It is a very small, low-mass device that minimizes the amount of implanted foreign material to reduce the risk of device-related complications. This article focuses on the anatomy of PFOs and compares the Coherex FlatStent™ with currently available and experimental PFO closure devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.