A potentiometric ion-selective electrode based on new compound, as a carrier, has been successfully developed for detection of perchlorate anion in aqueous solution. Within the perchlorate ion concentration range 1.0x10(-6) to 1.0 mol L(-1) the electrode had a linear response with a Nernstian slope of 60.6+/-1.0 mV per decade . The limit of detection as determined from the intersection of the extrapolated linear segments of the calibration plot was 8.0x10(-7) mol L(-1). The proposed electrode has fairly a good discriminating ability towards ClO(4) (-) ion in comparison to other anions. The sensor has a response time of < or =10 s and can be used for at least 2 months without substantial divergence in potential. It was successfully applied to direct determination of perchlorate in urine and water.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.