A potentiometric ion-selective electrode based on new compound, as a carrier, has been successfully developed for detection of perchlorate anion in aqueous solution. Within the perchlorate ion concentration range 1.0x10(-6) to 1.0 mol L(-1) the electrode had a linear response with a Nernstian slope of 60.6+/-1.0 mV per decade . The limit of detection as determined from the intersection of the extrapolated linear segments of the calibration plot was 8.0x10(-7) mol L(-1). The proposed electrode has fairly a good discriminating ability towards ClO(4) (-) ion in comparison to other anions. The sensor has a response time of < or =10 s and can be used for at least 2 months without substantial divergence in potential. It was successfully applied to direct determination of perchlorate in urine and water.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.