Abstract
The effect of Pd on analyte absorbance signals was evaluated for several elements including As, Cd, Cr, Pb, Sn and TI. Conventional furnace operating procedures and wall atomisation were used. The Pd reduction step was found to be essential for the recovery of Pb and TI when a halide matrix was present. Absorbance profiles were found to be sensitive to the mass of Pd present. Slight positive shifts in the peak temperature were noted for low (ng) levels of Pd, while higher levels of Pd caused increasing shifts in the peak temperature. Excluding Pb, the peak-area absorbances increased with Pd mass while peak-height absorbances generally reached a maximum at Pd masses below 2 µg. The data imply that the mass amount of the Pd chemical modifier should be considered during the development of analytical methods which include Pd.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.