Abstract
Preparative separations in supercritical fluid chromatography (SFC) involve the injection of large volumes of the solute. In SFC, the mobile phase is typically high pressure CO2+modifier and the solute to be injected is usually dissolved in the modifier. Two-types of injection methods, modifier-stream and mixed-stream, are common in commercial preparative SFC systems. In modifier-stream injection, the injection is made in the modifier stream which is later mixed with the CO2 stream, while in the mixed-stream injection, the injection is made in a mixed CO2+modifier stream. In this work a systematic experimental and modelling study of the two techniques is reported using single-enantiomers of flurbiprofen on Chiralpak AD-H with CO2+methanol as the mobile phase. While modifier-stream injection shows non-distorted peaks, mixed-stream injection results in severe peak-distortion. By comparing the modelling and experimental results, it is shown that the modifier “plug” introduced in the mixed-stream injection is the primary cause of the peak distortions. The experimental results also point to the possible existence of viscous fingering which contributes to further peak distortion.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.