Abstract

A novel peak diffusing measurement method is reported. In this method, a pulsed Gaussian heat source is focused on the material’s surface, and a thermal camera is utilized for detecting the temperature evolution. Specially, the peak temperature timetmaxat different distance from the heat source center is extracted using the Singular Spectrum Analysis (SSA). Thermal diffusivity is then calculated by fitting multiple sets of extractedtmaxinto the derived regression equation. The advantage of this method is that it does not require the high precision time synchronization between the excitation source and detector. Also, the calculation process is very simple, which achieves an efficient in situ measurement of the thermal diffusivity. The measured radial thermal diffusivity of samples prepared from Ti and Ni is in good agreement with the reference data with a 1.8% and a 3.9% error bound.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.