Abstract
Prospective ECG triggering has the potential of reducing radiation exposure while maintaining diagnostic accuracy of cardiac computed tomography (CT). The aim of this study is to review patient radiation doses associated with coronary artery calcium scoring (CACS) and CT coronary angiography (CTCA) and to compare results between prospective and retrospective acquisition schemes. Patient radiation doses from CACS and CTCA were extracted from 67 relevant studies. Mean effective dose for CACS and CTCA with prospective ECG triggering is significantly lower than retrospective acquisition, 0.9±0.4 vs. 3.1±1.4 mSv, p < 0.001, and 3.4±1.4 vs. 11.1±5.4 mSv, p < 0.001, respectively. In both cardiac CT examinations, application of dose modulation techniques result in significantly lower doses in retrospective schemes, however, even with dose modulation, retrospective acquisition is associated with significantly higher doses than prospective acquisition. The number of slices acquired per rotation and the number of X-ray sources of the CT scanner (single or dual source) do not have a significant effect on patient dose.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.