Abstract

to design a flow model to determine whether and why increased proximal neck angulation correlates with increased risk of proximal perigraft endoleak (PPE) flow. a tapered stent-graft (Gianturco stent + Dacron) was deployed with proximal and distal necks made of silicone. The amount of PPE was measured over 30-s periods while the angulation of the proximal neck was increased. PPE flow increased significantly for angulations > or =30 degrees. The graft was lifted off the neck wall and gaps between the stent-graft and the neck created. greater neck angulation increased PPE flow. The stent-graft was lifted off the neck and subsequent gaps created. This model could be used to compare available stent-grafts and to test new designs that would improve the seal in angulated necks.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.