Abstract

Abstract Context.—Previous studies have shown that paraproteins caused spurious results on individual analytes including total bilirubin (TBIL), direct bilirubin (DBIL), or HDL-cholesterol (HDL-C). Studies demonstrating paraprotein interferences with multiple analytes measured by different analyzers have not been reported. Objective.—To systemically investigate interferences of paraproteins on TBIL, DBIL, and HDL-C measured by the Roche MODULAR and the Olympus AU2700. Design.—Eighty-eight serum specimens with monoclonal gammopathies were analyzed using the Roche MODULAR and the Olympus AU2700. Paraprotein interferences with the MODULAR and AU2700 were identified by abnormal absorbance curves and confirmed by results from the Ortho Vitros 950 or inconsistent laboratory information. Results.—Spurious results occurred in 89 of 528 measurements; 29 specimens did not demonstrate any interferences whereas 26 specimens gave spurious results in 2 to 4 of the 6 assays. Paraprotein interferences caused spuriously h...

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.