Abstract

Coincidence timing calibration is fundamental to PET imaging. The electronics, cable lengths, and detector physics such as charge drift and depth dependence add to the measured time differences in coincidence sorting - increasing random rate, decreasing true rate, and degrading system performance. This work investigates the parameter selection for convex optimization (Ordinary Least Squares) for timing calibration. We test the correlation between commonly selected parameters and the experimentally measured coincidence time difference. Additionally, we test 127 nested models of a parameterized regression equation to identify the those which optimize MSE, BIC, and FWHM, respectively. In each of these models, the FWHM performance improved ~53%, though the value shifted from ~ 304 to 160 ns - far from ~ 10 ns FWHM CZT can achieve. The results point to the lack of a necessary parameter, such as trigger threshold level or temperature, or data which is too variable for the OLS optimization.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.