Abstract
We have developed a new technique for the detection of neutrons in the presence of a high gamma-ray background. This was accomplished by using two lithium-loaded glass scintillators (/sup 6/Li//sup 7/Li) in combination. The neutron capture reaction in /sup 6/Li has a very high Q-value which means better discrimination against gamma-ray background. The /sup 7/Li detector, on the other hand, is neutron insensitive and was used to measure separately the gamma contribution in a mixed neutron-gamma field. By subtracting this gamma contribution from the /sup 6/Li data, which is sensitive to both gamma rays and neutrons, one can obtain the pure neutron counts.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.