Abstract

The majority of Dutch F-16 pilots experience in-flight oxygen mask related nasal discomfort and injury. We aimed to analyze the fit of the oxygen mask. We successfully scanned 35 pilots with a 3D scanner to measure the distance between the contact area on the nose and the oxygen mask in seven points, and the difference (delta) in angle between the nose and the mask relative to the facial midline. Ill fit was defined as > or = 4 mm distance, or delta > or = 4 degree angle. Intraclass correlation coefficients were used to objectify the reproducibility of the measurements. There were 35 pilots who had > or = 4 mm distance between the nasal dorsum and the mask and 14 pilots had delta > or = 4 degree angle. Reproducibility of the distance measurements was excellent (R > or = 0.93, CI 95% 0.86-0.96) and of the angle measurements was good (R = 0.72, CI 95% 0.5-0.8). The room for displacement that we found may be a contributor to the discomfort. It is indicative of an ill fit of the oxygen mask, since all subjects had complaints. We recommend evaluating whether improved fixation of the mask and simultaneous reduction of the pressure on the nose will improve comfort.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.