Abstract

Abstract Ti-13Nb-13Zr (Ti-13-13TM) is a new titanium alloy developed for medical implant applications. In the oxygen diffusion hardened condition, Ti-13-13 and titanium alloys which possess zirconium, are hardenable at temperatures of 400 to 600°C with the end result being a passive, mechanically stable, abrasion resistant oxide ceramic surface layer of about 0.2 to 2 μm in thickness. The thickness of this surface layer depends on the concentration of zirconium, time, temperature, and partial pressure of oxygen. This paper describes the results of x-ray diffraction analysis, secondary ion mass spectroscopy, x-ray photoelectron spectroscopy, Knoop and Nano hardness measurements, and reciprocal polarization resistance measurements of hardened Ti-13-13, and selected Ti-Nb-Zr alloys. The results of this work suggest that zirconium plays a key role for effective oxygen diffusion hardening at 500°C for Ti- ] 3-13 and similar Ti-Nb-Zr alloys.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.