Abstract
A practical and sensitive method to assess volatile organic compounds (VOCs) from JP-8 jet fuel in human whole blood was developed by modifying previously established liquid–liquid extraction procedures, optimizing extraction times, solvent volume, specific sample processing techniques, and a new on-column large-volume injection method for GC–MS analysis. With the optimized methods, the extraction efficiency was improved by 4.3 to 20.1 times and the detection sensitivity increased up to 660 times over the standard method. Typical detection limits in the parts-per-trillion (ppt) level range were achieved for all monitored JP-8 constituents; this is sufficient for assessing human fuels exposures at trace environmental levels as well as occupational exposure levels. The sample extractions are performed in the field and only solvent extracts need to be shipped to the laboratory. The method is implemented with standard biological laboratory equipment and a modest bench-top GC–MS system.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Chromatography B: Biomedical Sciences and Applications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.