Abstract

This study presents an optimization of the detection coil of high-Tc superconducting quantum interference device (SQUID)-based nuclear magnetic resonance (NMR) in microtesla fields for discriminating a minimum amount of liver tumor in rats by characterizing the longitudinal relaxation rate, T1−1, of tested samples. The detection coil, which was coupled to the SQUID through a flux transformer, was optimized by varying the copper wires’ winding turns and diameters. When comparing the measured NMR signals, we found that the simulated NMR signal agrees with simulated signals. When discriminating liver tumors in rats, the averaged longitudinal relaxation rate was observed to be T1−1 = 3.3 s−1 for cancerous liver tissue and T1−1 = 6.6 s−1 for normal liver tissue. The results suggest that it can be used to successfully discriminate cancerous liver tissue from normal liver tissues in rats. The minimum amount of samples that can be detected is 0.2 g for liver tumor and 0.4 g for normal liver tissue in 100 μT fields. The specimen was not damaged; it can be used for other pathological analyses. The proposed method provides more possibilities for examining undersized specimens.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.