Abstract

A computer-assisted retention prediction system which is an alternative to computer simulation is demonstrated to be an efficient tool for finding optimum gradients in anion-exchange separations of oligonucleotides. The effect of the gradient profile on the resolution is illustrated by computer simulation in the visualization of simulated chromatograms by changing the variables for the gradient profile. A convex gradient was found to be more effective for the separation of high-molecular-weight oligonucleotidas than a linear gradient which has been widely used. An increase in the gradient time also improved the resolution. Recommendations are presented for an approach to maximize the resolution and to minimize the analysis time in the anion-exchange high-performance liquid chromatography of olignonucleotides.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.