Abstract

Near-infrared spectroscopy (NIRS) has been proposed as a noninvasive modality for detecting complications in patients undergoing extracorporeal membrane oxygenation (ECMO), and it can simultaneously reveal the global circulatory status of these patients. We optimized ECMO therapy on the basis of real-time peripheral NIRS probing. Three patients underwent venoarterial (VA) ECMO and one patient underwent venovenous (VV) ECMO. All patients received peripheral ECMO cannulation with routine distal perfusion catheter placement. We designed an experimental protocol to adjust ECMO blood flow over 1 hour. Hemodynamic responses were measured using NIRS devices attached to the calf at approximately 60% of the distance from the ankle to the knee. HbO2 levels change substantially with adjustments in ECMO flow, and they are more sensitive than HHb levels and the tissue saturation index (TSI) are. NIRS for optimizing ECMO therapy may be reliable for monitoring global circulatory status.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.