Abstract

In this work, a numerical model was developed for critical evaluation of the 13CO2/12CO2 ratio retrievals ( value) from laser absorption spectra. The goal of the analysis was to determine the dependency of the absolute error of δ on different experimental parameters, in order to find the optimal conditions for isotopic ratio retrievals without using calibrated reference samples. In our study, the target precision for was set at a level of %. The analysis was performed in the spectral range of the CO2 band at 1.6 μm, with the theoretical data originating from the HITRAN database. The proposed fitting algorithm allowed efficient compensation of the interference from weak transitions which are not well recognizable in a single spectrum. This effect was found to make a dominant contribution to the value. Next, the optimal conditions for such an experiment regarding the pressure, spectral range and spectrum noise were found and discussed from the perspective of widely tunable laser applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.