Abstract
A strategy was developed for the control, standardization, and critical evaluation of an enzyme-linked immunosorbent assay (ELISA) for the detection of human papillomavirus-specific immunoglobulin G in human sera. Control human sera, polyclonal animal sera, and monoclonal antibodies were used to establish optimal assay parameters, including antigen coating, serum dilutions, and criteria for daily reproducibility, monitoring, and rejection of assays. Three evaluation techniques were used in parallel to define an optimal cutoff absorbance value that yields greater than 93% sensitivity and 98.5% specificity in the assay's ability to discriminate positive and negative control sera. This strategy provides an optimal method by which to determine cutoff absorbance values for ELISA.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.