Abstract
Capillary zone electrophoresis was optimized to quantitatively determine codeine and paracetamol via central composite factorial design. Critical parameters (concentration, buffer, pH, voltage) assessed effects on resolution, analysis time and efficiencies. Optimum separation conditions were achieved using phosphate buffer 20 mM (pH 6.8) and voltage (15 kV). The optimized procedure easily determined codeine and paracetamol with separation in less than 3 min. Calibration curves (R > 0.999) were prepared, with LODs of 13.5 and 340 ng mL(-1) for codeine and paracetamol, respectively, and a good R.S.D.% (<3%). This method was applied to determine codeine and paracetamol in pharmaceutical formulations; recoveries coincided with stated contents.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.