Abstract
Optimal beam geometry for kilovoltage cone-beam tomosynthesis (CB-TS) imaging for the set-up of 20 patients with breast cancer was analyzed and presented for prone and supine scenarios. Gantry angles and field sizes that allow maximum clearance and minimum dose to normal organs from kilovoltage imaging were determined for medial and lateral arcs. Optimal CB-TS arc span may vary depending on patient geometry, particularly on the separation between the cavity and the chest wall. CB-TS arc angles ⩾40° were found to be optimal for the cavity to chest wall separation d > 3 cm for supine set-up and d > 7 cm for prone set-up. Maximum required jaw openings were 20 × 14 cm 2. Recommendations are provided for clinical implementation of CB-TS imaging for breast cancer patients.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.