Abstract

Protective glasses from the back of modern smartphones were tested as potential emergency dosimeters using optically stimulated luminescence (OSL). The glasses were shown to exhibit several favorable dosimetric properties including a strong sensitivity to radiation, a linear dose-response relationship from 0.1 to 8 ​Gy, a 3% reproducibility, and a minimum detectable dose in the low tens of milligray, even after significant fading. The OSL signal was observed to lose ~78% and ~90% of the original intensity over 1 ​h and 1 week, respectively; however, a fading curve was developed to correct for any such signal loss. A dose-recovery test demonstrated that reconstructed doses were capable of accuracies within 20%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.