For high precision applications of optical microscopes, it is critical to achieve symmetrical angular illumination intensity at the sample plane, in addition to uniform spatial irradiance achieved by Köhler illumination. A correlation between the angular illumination asymmetry and the contrast in the image of a line grating target was demonstrated as the target is scanned through focus. Using this correlation, we present a novel, yet simple method of experimentally evaluating the angular illumination asymmetry (ANILAS) at the sample plane of an optical microscope across the field of view. This ANILAS map is expected to be a useful method for assessing the illumination condition of optical systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.