Abstract

Variable results of strabismus surgery may be due in part to errors in prism measurement. The amount of deviation neutralized by an ophthalmic prism is variable depending on how the prism is held. For example, a 40Δ glass prism with the posterior face held in the frontal plane gives only 32Δ of effect. Glass prisms are calibrated for use in the Prentice position. Plastic prisms are calibrated for use in the frontal plane position. Surprisingly large errors in prism measurement are produced when adding a small prism to a large prism. For example, adding a 5Δ glass prism to a 40Δ glass prism gives not 45Δ of effect, but 59Δ. This error can be minimized but not eliminated by holding one prism in front of each eye. The error can also be calculated so that the appropriate correction can be made.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.