Abstract
The Visible Infrared Imaging Radiometer Suite (VIIRS) instrument is a whiskbroom system with 22 spectral bands split between 16 moderate resolution bands (M-bands), five imagery resolution bands (I-bands) and a panchromatic day-night band. Latitude and Longitude geolocation data are generated for each pixel at the M-band, I-band and day-night band spatial resolutions based upon various instrument parameters including focal length. In this study we measure the focal length of the VIIRS instrument from on-orbit data. This is achieved by simulating VIIRS band I2 using Landsat 8 TM band 5 utilizing the VIIRS instrument system point spread function (PSF) and geolocation data generated with varying values of focal length. The focal length value that produces the highest spatial correlation between the original and simulated VIIRS data is taken to be the measured instrument focal length.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.