Abstract

The application of coupled-column liquid chromatographic analysis to pharmacokinetic studies eliminates the need for sample clean-up from plasma. Considering lipophilic antineoplastic agents, we tested this approach to analyze paclitaxel under unfavourable circumstances (i.e., weekly low-dose regimen, plasma protein binding >90%, UV detection at 229 nm). The excellent quality control data (recovery: 95.6–100.7%, inter-assay relative standard deviation on 5 days: 1.3–3.2%, accuracy: 0.9–2.7%) and the detection limit of 19 n M indicates the usefulness of this method for the analysis of paclitaxel in plasma using on-line solid-phase extraction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.