Abstract

Nanoelectrospray ionization (nESI) coupled online with high-field asymmetric waveform ion mobility spectrometry (FAIMS) for small molecule analysis in a discovery pharmaceutical setting was examined. A conventional capillary pump, autosampler and nESI source were used to introduce samples directly into the FAIMS device. The FAIMS device was used to separate gas-phase ions on a timescale that was compatible with the mass spectrometer. The capability of the nESI-FAIMS combination to efficiently remove metabolite interferences from the parent drug, and reduce ion suppression effects, was demonstrated. On average, 85% of the signal intensity obtained from a neat sample was preserved in the extracted plasma samples. Standard curves were prepared for several compounds. Linearity was obtained over approximately 3 to 4 orders of magnitude. Comparison of results from nESI-FAIMS with those from conventional LC/MS for a mouse pharmacokinetic study yielded concentration values differing by no more than 30%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.