Abstract

Calculation of lead concentration from K-series X-ray fluorescent studies uses a robust normalization technique based on the amplitude or area of the elastic signal. Parameter estimation of the elastic signal can be affected by the overlap of the K β2 line, especially for concentrations greater than 40 ppm where the K β2 amplitude can be greater than 1% of the elastic signal. We tested the combination of estimation by method of least moduli and doublet deconvolution. We found that the estimation of the area of the elastic signal is more robust to changes in the low-energy end of the region of interest with the combined method than with method of least-squares estimation and singlet processing. We recommend use of the combined method for creation of calibration curves at concentrations greater than or equal to 40 ppm.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.