Abstract
Modern commercial direct-current current transformers (DCCT) can measure currents up to the kA range with accuracies better than 1E-5. We discuss here a DCCT calibration method and its implementation with commercial instruments typically employed in low resistance calibration laboratories. The primary current ranges up to 2 kA; in the current range below \SI{100}{\ampere} the calibration uncertainty is better than 3E-7. An example of calibration of a high-performance DCCT specified for primary currents measurement up to 900 A is discussed in detail.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.