Abstract

Focal distance is a key parameter for a focused ultrasound probe, especially in mouse brain stimulation where targets are right below the skull. A closed-form solution for the minimal focal distance with a given transducer size was derived in this study to facilitate precise focal spot alignment with targets in the mouse brain. The spherical profile corresponding to the minimal focal distance does not produce accurate focusing. An iterative algorithm based on Snell's law was introduced for lens profile calculation. With a suitable step size, an accurate lens profile can be obtained for the minimal focal distance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.