Abstract
An integrated system for DNA sequencing based on a nanoreactor for cycle-sequencing reaction coupled with on-line capillary zone electrophoresis (CZE) for purification and capillary gel electrophoresis (CGE) for separation is presented. Less than 100 nl of premixed reagent solution, which includes dye-labeled terminator pre-mix, bovine serum albumin and template, was hydrodynamically injected into a fused-silica capillary (75 μm I.D.) inside a laboratory-made microthermocycler for cycle sequencing reaction. In the same capillary, the reaction products were purified by CZE followed by on-line injection of the DNA fragments into another capillary for CGE. Over 540 base pairs (bp) of DNA can be separated and the bases called for single-standed DNA with 0.9% error rate. The total time was about 3.5 h, or a cycle time of 2 h with staggered operation. For double-stranded DNA, a longer reaction time was required and base calling up to 490 bp with 1.2% error rate was achieved. The whole system is readily adaptable to automated multiplex operation for DNA sequencing or polymerase chain reaction analysis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.