Abstract

An integrated system for DNA sequencing based on a nanoreactor for cycle-sequencing reaction coupled with on-line capillary zone electrophoresis (CZE) for purification and capillary gel electrophoresis (CGE) for separation is presented. Less than 100 nl of premixed reagent solution, which includes dye-labeled terminator pre-mix, bovine serum albumin and template, was hydrodynamically injected into a fused-silica capillary (75 μm I.D.) inside a laboratory-made microthermocycler for cycle sequencing reaction. In the same capillary, the reaction products were purified by CZE followed by on-line injection of the DNA fragments into another capillary for CGE. Over 540 base pairs (bp) of DNA can be separated and the bases called for single-standed DNA with 0.9% error rate. The total time was about 3.5 h, or a cycle time of 2 h with staggered operation. For double-stranded DNA, a longer reaction time was required and base calling up to 490 bp with 1.2% error rate was achieved. The whole system is readily adaptable to automated multiplex operation for DNA sequencing or polymerase chain reaction analysis.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.