Abstract

Chronic radiation exposure increases the risk of skin damage of medical personnel engaged in radiology. However, hand dose measurements in computed tomography (CT) for diagnostic purposes have not been evaluated. The occupational radiation dose to the hands of CT assistants was herein investigated to evaluate its compliance with the equivalent dose limit for the hand (500mSv/year). The occupational doses of nine CT assistants were measured in 89 cases (April 2017-May 2018) by installing radio-photoluminescence glass dosemeters (GD-302M) (70-μm dose-equivalent conversion coefficient=0.37) on the dorsal aspect of both hands. The occupational dose to the hand was the highest with head holding (right: 1.14mSv/CT scan, left: 1.07mSv/CT scan). Considering the results for annual work, even for head holding, the hand dose of the CT-assisting personnel was insignificant. However, CT assistants should be mindful of the possibility of locally higher doses to hands.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.