Abstract

We observed anomalous instrument responses above certain voltages when characterizing aggregates of airborne carbon nanotubes or nanofibers using a differential mobility analyzer (DMA). These were associated with sudden increases in measured number concentrations at high voltages, fluctuations in DMA voltage and audible high-frequency sounds from the DMA column. Onset of the anomalies depended on the material forming the aerosol and DMA sampling flow rate. The low density of the nanotubes relative to the nanofibers is thought to be the main reason the anomalous responses occur more easily and strongly with the nanotubes. Two possible mechanisms are suggested to explain the observations. The results indicate that measurement of nanometer-diameter conducting fibrous material by electrical mobility analysis may present a unique challenge.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.