This paper discusses some issues regarding the structural behavior of a Balloon-Expandable (BE) stent made of stainless steel material (AISI316L). BE stent is a tubular, often mesh-like, structure which is expanded inside a diseased (stenosed) artery segment to restore blood flow and keep the vessel open following angioplasty. Most of BE stent designs have two fundamental constituents: expandable ring elements, and connecting elements “bridges” which connect adjacent rings together. The stent design is a major factor which determines its reliability during insertion into the blocked artery and throughout the long term in vivo service. The objective of this paper is to study the structural behavior of the stent in order to show the effect of the stent design, especially the geometry of the “bridges” in terms of flexibility, torsion and expansion. The numerical investigation is based on Finite Element Analysis (FEA) using Abaqus© finite element code. It has been demonstrated by FEA that the geometry of the connecting elements “bridges” has a significant impact on the structural behavior of the stent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.