Abstract
A novel sealed gaseous parallel plate avalanche chamber detector is described which is significantly less prone to discharges and can consequently achieve high gas gains at high counting rates. The detector has demonstrated stable gains greater than 104 at counting rates in excess of 107 counts/mm2 s.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.