Abstract
MPL exon 10 mutations were the second class of mutations shown to be associated with the pathogenesis of some Philadelphia chromosome – negative myeloproliferative neoplasms (MPNs). Recently, their identification gained wide recognition in the diagnostic work-up for suspected cases of JAK2 V617F negative MPNs. Various molecular approaches have been applied, yet universally accepted method is still lacking. We aimed at development and validation of a novel bead-based liquid assay using Locked nucleic acids (LNA)-modified oligonucleotide probes for multiplexed detection of the following MPL mutations: W515L/K/A/R. Testing on both artificial plasmid constructs and on clinical samples revealed that the method was comparable in terms of specificity to direct sequencing and had a much higher sensitivity of 1% mutant alleles. This method could be successfully implemented in the diagnostic work-up for MPNs. Furthermore, this system allows further multiplexing for single-tube identification of different mutations associated with MPNs.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.