Abstract

We herein describe a novel device to contain droplets and aerosols during phacoemulsification. We modified the silicon phaco test chamber into an aerosol containment chamber (ACC) by shortening the chamber and making a pear-shaped opening at one aspect of its tip. The ACC was fitted over phaco tip such that 4-5 mm of phaco tip and sleeve was exposed. When the phaco tip and irrigation port are inside the anterior chamber during phacoemulsification, the portion of the modified chamber remains around the clear corneal tunnel in an enclosing manner that contains aerosols and droplets.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.