Abstract

This preclinical study emulating the clinical environment quantitatively analysed the accuracy of pedicle screw insertion using a navigated robotic system. Pedicle screws were placed from T7 to L5 in the whole-body form of a cadaver. After the insertion of multiple artificial markers into each vertebra, errors between the planned insertion path and the inserted screw were quantified using the Gertzbein-Robbins system (GRS) and offset calculation. A total of 22 screws were placed. Almost all (95.45% [21/22]) were classified as GRS A or B, while one (4.55%) was GRS C. The mean and standard deviations of entry, tip, and angular offset were 1.78±0.94mm, 2.30±1.01mm, and 2.64±1.05°, respectively. This study demonstrated that pedicle screw insertion using a navigated robotic system had high accuracy and safety. A future clinical study is necessary to validate our findings.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.