Abstract
Herein, an electrochemical sensor based 5-(2-phenylbenzo[b]thiophen-3-yl)furan-2-carbaldehyde (PTFC) was developed for the detection of cancer antigen 125 (CA125). Measurements were obtained via cyclic voltammetry (CV), electrochemical impedance spectroscopy (EIS), and differential pulse voltammetry (DPV), and square wave voltammetry (SWV) techniques to investigate features of the electrochemical sensor such as concentration effect of CA125, incubation time, scan rate, limit of quantification (LOQ), limit of detection (LOD), and interference effect of structures found in serum. The sensor was found to have a fairly wide linear range as 0.01–100 ng mL−1 and 100–1000 ng mL−1. LOQ and LOD values were determined as 0.024015 ng mL−1 and 0.008005 ng mL−1 (S/N = 3), respectively. These results showed that the PTFC-based electrode could be a promising electrode for the detection of CA125.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.