Abstract

Long-bone fractures are very common in trauma centers. The conventional Arbeitsgemeindschaft fur Osteosynthesefragen (AO) technique contributes to most fracture healing problems, and external fixation technology also has several disadvantages, so new techniques are being explored. A novel hexapod computer-assisted fracture reduction system based on a 3D-CT image reconstruction process is presented for closed reduction of long-bone diaphyseal fractures. A new reduction technique and upgraded reduction device are described and the whole system has been validated. Ten bovine femoral fracture models were used with random fracture patterns. Tests results were as follows: residual deviation 1.24 + 0.65 mm for the axial deflection, 1.19 + 0.37 mm for the translation, 2.34 + 1.79° for the angulation, and 2.83 + 0.9° for the rotation. The reduction mechanism has the advantages of high positioning, reduction and computer accuracy, and intra-operative stability for both patients and surgical team. With further investigation, it could be applied in many kinds of long-bone diaphyseal fractures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.