Abstract

A novel flow-injection chemiluminescence(CL) system has been established based on the inhibitory effect of nonylphenol(NP) on the morin-Ce4+ CL system. The flow path, flow rate, concentration of Ce4+, pH of Ce4+ solution and concentration of morin were discussed. The results show that the best determination conditions are 5.0×10-3 mol/L of Ce4+, 1.0×104 mol/L of morin and pH=1.2 of Ce4+ solution. Under the optimal conditions, the reduction value of CL intensity is linear with the concentration of NP within 2×10-8~2×10-7g/ml and 2×10-6~4×10-5 g/ml. The regression equation was Δl1=12.12+225.95CNP and Δl2=65.291+753164.6CNP with correlation coefficient 0.9995 and 0.998, respectively. And the detection limit was 2×10-10 g/ml (3σ). The relative standard deviation (RSD) is 1.48% for the determination of 2×107 g/ml NP(n=10). The recoveries of NP are between 90.7%~106.3% for three different environmental water samples. It suggested that this method can be used for the determination of NP in the real environmental samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.