Abstract
A timely diagnosis of vascular compromise improves free flap survival. The aim of this study was to determine the feasibility of monitoring flap vascular patency through noninvasive microcirculation monitoring using SDF. An experimental study was performed in five volunteers. A radial forearm flap was simulated in both upper extremities and cutaneous microcirculation assessment devices were used (SDF and NIRS), and compared with clinical and doppler evaluation. Conditions of venous occlusion (VO) and total vascular occlusion (TVO) were mimicked by using a cuff. The results of the Microscan SDF monitoring during VO and TVO showed a reduced microcirculatory flow at 16.5 (6-30) and 6 (2-11) seconds respectively. Both NIRS and clinical evaluation were slower than SDF at identifying vascular compromise. During VO the Doppler signal does not disappear; while in TVO the signal disappeared at 1.8 (1-5) seconds. This study shows that evaluation of microcirculation with Microscan SDF is a viable alternative that may allow detection of flap venous and total vascular occlusion earlier than Doppler, NIRS and clinical evaluation
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.