Abstract

A deadbeat controller has been proposed for the control of pulsatile pump flow in an implantable rotary blood pump (IRBP). A lumped parameter model of the cardiovascular system, in combination with the stable dynamical models of pulsatile flow and differential pressure (head) estimation for the IRBP was used to evaluate the controller. Pump speed and current were used as the only measured variables of the control system. The control algorithm was tested using both constant and sinusoidal reference pump flow input, under healthy and heart failure conditions. Results showed that the controller is able to track the reference input with minimal error in the presence of model uncertainty.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.