Abstract

This study evaluated the feasibility of employing non-invasive intra-cardiac pressure estimation using subharmonic signals from ultrasound contrast agents in humans. This institutional review board–approved proof-of-concept study included 15 consenting patients scheduled for left and right heart catheterization. During the catheterization procedure, Definity was infused intra-venously at 4–10 mL/min. Ultrasound scanning was performed with a Sonix RP using pulse inversion, three incident acoustic output levels and 2.5-MHz transmit frequency. Radiofrequency data were processed and subharmonic amplitudes were compared with the pressure catheter data. The correlation coefficient between subharmonic signals and pressure catheter data ranged from −0.3 to −0.9. For acquisitions with optimum acoustic output, pressure errors between the subharmonic technique and catheter were as low as 2.6 mmHg. However, automatically determining optimum acoustic output during scanning for each patient remains to be addressed before clinical applicability can be decided.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.