Abstract

Inefficacy of implantable cardioverter/defibrillator (ICD) shocks or inappropriate shocks in ICD patients may occur due to mechanical or technical failure of the device. In our unusual case of ICD defibrillation failure, faulty insertion and configuration of the defibrillation lead in the header port resulted in ineffective shock therapies and a non-determinable defibrillation threshold. In cases of successless defibrillations during or after ICD implantation, this problem should be considered.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.