A simple noise thermometer, based on the measurement of the currents produced by the noise voltages over a 44 μω manganin resistor with a commercial SQUID, has been used for temperature measurements between 0.1 K and 0.4 K. Results at the superconductive transition points of AuIn 2, AuAl 2 and Ir (nominally at 208 mK, 162 mK and 100 mK), realized with a NIST SRM 768 device, are consistent with the assigned NIST temperature values within 0.15 percent. 3He melting pressures have been measured at these three points and at the minimum of the melting curve, ( p min=(2.9311 4±0.0002)MPa). Existing data for the melting curve are compared with results of thermodynamic calculations between 2 mK and 250 mK.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.